Home / ECG leadwires and trunk cables, matched / ECG trunk cables — the monitor-side half of

ECG trunk cables — the monitor-side half of the ECG pathway

Listed trunks24
PlatformsGE · Philips · Mindray…
Known refs2106308-001…
Mates with3/5/6/10-lead sets

The trunk cable is the fixed half of the ECG pathway: it lives on the monitor, takes the knocks of daily bed moves, and outlasts several leadwire sets — until it doesn't, and every lead position starts misbehaving at once. The trunk range (24 listed variants) covers the mainstream platforms, including GE Marquette host cables cross-referenced to OEM numbers such as 2106308-001 (12-lead trunk) and 2017003-001 (3/5-lead trunk), plus Philips, Mindray, Nihon Kohden and Edan patterns.

Trunks are specified by the pair they join: monitor-side connector and generation on one end, the leadwire-set socket (and its pin/keying pattern) on the other. A trunk from the wrong generation may share a monitor brand and still not mate with your sets — which is why this page family exists separately from the sets themselves.

Philips-compatible ECG trunk cable
ECG trunk cable, Philips-compatible — official store product photo

Diagnosing a trunk fault

Suspect the trunk when artefact or lead-off alarms appear across multiple lead positions, when a known-good leadwire set does not cure the fault, or when the fault moves with cable handling near the monitor end. Trunk conductors fatigue exactly where they flex — at the monitor connector and at the set socket. A wiggle test under monitoring (watch the tracing while flexing each segment gently) localises the break in under a minute.

Ordering the right trunk

Give the monitor model and, if the old trunk carries one, its OEM part number — GE trunks in particular span several connector generations under similar names, and the number settles it. State also the leadwire-set family the trunk must accept (3/5-lead monitoring or 10-lead diagnostic), since socket patterns differ.

Representative products from the current catalogue

Product (official store listing)Key specifications
12 Lead ECG Trunk Cable GE Marquette Compatible 2106308-001AMC: CB-5110006;GE Healthcare > Marquette: 2017006-001 (12 ft), 416035-001, 416035-003, 2106308-001 (12 ft), 2017006-002 (4 ft), 2106308-002 (4 ft);Sage: T02-12-30M;Datex Ohmeda: E-PRESTN, M-PRESTN;GE Healthcare > Critikon > Dinamap: CARESC
2016560-001 GE Host Cable (Date Cable) Compatible for Cam 14 MAC500,MAC5500AMC: CB-2016560-001;GE Healthcare > Marquette: 2016560-001, 700657-001, E9001YT, 2104403-001 (5.5 ft);Pacific Medical: NEGE9001;Sage: V02-13M;Category: EKG;Monitor End Connector: Flat, 10-pin Connector;Total Cable Length: 5.6ft (1.7m);Cable
2017003-001 ECG Trunk Cable GE Marquette Compatible 3/5 Lead-IECAMC: CB-515006 (10 ft Cable);Curbell: CBT-05ND-10CM-0001;GE Healthcare > Marquette: 2017003-001 (12 ft), 2017003-002 (4 ft), 2022948-002, 412931-001, E9002ZB, 164L0025, 2106305-001 (12 ft), 2106305-002 (4 ft), 2106311-001 (20 ft);GE Healthc
2104723-001 GE HealthCare ECG Trunk Cable Compatible With 10 Leads-22341809GE Healthcare > Marquette: 22341809, 2104723-001;Sage: T02-10-25M;Tenacore: 45509-B;Category: EKG;Monitor End Connector: DB-15 Connector;Total Cable Length: 7.2ft (2.2m);Cable Material: TPU Jacket;Lead Color Coding: AHA/IEC;Lead Cable Color
6 Lead ECG Trunk Cable GE Marquette Compatible 2106307-001Datex Ohmeda: Aestiva/5, E-PRESTN, M-PRESTN, Unity;Draeger: Narkomed 6400;GE Healthcare > Critikon > Dinamap: Carescape B105, Carescape B125, Carescape B450, Carescape B650, Carescape B850, ProCare B20, Procare B40;GE Healthcare > Marquette
Compatible ECG Trunk Cable EC419-4IECG, ECG Trunk Cables
Draeger Compatible ECG Trunk Cable - 3368391Draeger: 3368391, 5950192 (1.5 meter cable), 5950196;Sage: T09-25M;Category: ECG;Monitor End Connector: Rectangular, 16-pin Draeger Infinity Connector;Total Cable Length: 8.2ft (2.5m);Cable Material: TPU Jacket;Cable Diameter: 4mm;Lead Colo
ECG Trunk Cable Mindray Compatible for Neonatal DIN 3-LeadMindray: EV6222,040-000754-00;Category: ECG;Monitor End Connector: 12-pin Connector;Total Cable Length: 9ft (2.7m);Cable Material: TPU Jacket;Cable Diameter: 5mm;Lead Color Coding: AHA/IEC;Lead Cable Color: Gray;Packaging Unit: 1;Weight: 0.

Questions & answers

Will a trunk cable from the same brand always fit my sets?

No — connector generations differ within a brand. Match by monitor model and OEM trunk reference, which we confirm before shipping.

Can a trunk cable be repaired?

Trunks are replaced, not repaired: conductor fatigue is distributed, and a repaired trunk is an intermittent fault waiting for a night shift.

Do you stock 12-lead trunks?

Yes — 10-lead diagnostic trunks (for 12-lead acquisition) are listed for the major ECG platforms, including GE Marquette references.

See also: 3/5-lead ECG cables · ECG lead placement guide · All ECG cables

Request a compatibility check & quotation

Tell us the monitor brand and model — or the OEM part number of the accessory you are replacing — and the match is confirmed before pricing.